Thị Thanh Phương N, Thị Thanh Đông Đỗ, Thị Ngọc Hạnh H, Trí Dũng N, Thị Minh Hương V. SUMMARY MONITORING SIX SIGMA AND QGI IN THE EVALUATION AND FOLLOW-UP OF QUALITY FOR SELECTED BIOCHEMICAL TESTS ON THE AU5800 ANALYZER SYSTEM. JHA [Internet]. Vietnam; 2026 Jun. 7 [cited 2026 Jun. 18];2(8):141–146. https://tcsuckhoelaohoa.vn/bvtn/article/view/163 doi: 10.63947/bvtn.v2i8.20
Loading...
Loading citation...

Abstract

To apply Six Sigma and Quality Goal Index (QGI) for evaluating and improving the quality of 10 routine biochemical tests (glucose, urea, creatinine, uric acid, ALT, GGT, calcium, magnesium, iron, CK) on the AU5800 analyzer at the Biochemistry Department, Thong Nhat Hospital. Subject and methods: he study utilized internal quality control (IQC) data from the AU5800 system at two control levels (Level 2 and Level 3) across two phases. Parameters such as bias, coefficient of variation (CV%), and Six Sigma metrics were calculated based on total allowable error (TEa), combined with analysis of the Quality Goal Index (QGI). Results: In Phase 1, 6/10 tests achieved a Sigma ≥4, while Urea and Creatinine had a Sigma of 2 with low QGI. In Phase 2, Sigma values for Urea and Creatinine improved to ≥ 3. Conclusion: The combination of Six Sigma and QGI enables early detection of errors, effective targeted improvement — especially for urea and creatinine — and demonstrates superior value and sustainability in the quality management of biochemical tests on the AU5800 analyzer.

Keywords

Six Sigma QGI Au5800

References

  1. Westgard JO. Six sigma quality design and control: Westgard QC, Incorporated Madison; 2001.
  2. Westgard JO. Useful measures and models for analytical quality management in medical laboratories. Clinical Chemistry and Laboratory Medicine (CCLM). 2016;54(2):223-33. DOI: https://doi.org/10.1515/cclm-2015-0710
  3. Verma M, Dahiya K, Ghalaut VS, Dhupper V. Assessment of quality control system by sigma metrics and quality goal index ratio: A roadmap towards preparation for NABL. World journal of methodology. 2018;8(3):44. DOI: https://doi.org/10.5662/wjm.v8.i3.44
  4. Zhou B, Wu Y, He H, Li C, Tan L, Cao Y. Practical application of Six Sigma management in analytical biochemistry processes in clinical settings. Journal of clinical laboratory analysis. 2020;34(1):e23126. DOI: https://doi.org/10.1002/jcla.23126
  5. Guo X, Zhang T, Gao X, Li P, You T, Wu Q, et al. Sigma metrics for assessing the analytical quality of clinical chemistry assays: a comparison of two approaches: Electronic supplementary material available online for this article. Biochem Med (Zagreb). 2018;28(2):020708. doi: 10.11613/bm.2018.020708. PubMed PMID: 30022883; PubMed Central PMCID: PMC6039159. DOI: https://doi.org/10.11613/BM.2018.020708
  6. Mao X, Shao J, Zhang B, Wang Y. Evaluating analytical quality in clinical biochemistry laboratory using Six Sigma. Biochemia medica. 2018;28(2):253-6. DOI: https://doi.org/10.11613/BM.2018.020904

License

© 2026 The Author(s). Published by Journal of Health and Aging.