APPLICATION OF SIX SIGMA IN THE EVALUATION AND IMPROVEMENT OF BIOCHEMISTRY TESTING PERFORMANCE AT THONG NHAT HOSPITAL

Đỗ Thị Thanh Đông 1 , , Nguyễn Thị Nhạn 2 , Ngô Kiến Đức 1, 3 , Lê Hoàng Lâm 1 , Nguyễn Thị San 1 , Lê Đình Thanh 1
1 Thong Nhat Hospital image/svg+xml
2 Trường ĐH Nguyễn Tất Thành image/svg+xml
3 Ho Chi Minh City Medicine and Pharmacy University image/svg+xml
* Corresponding author:

Article Information

Metrics
Downloads: 4 Views: 12
Published
2025-07-10
Section
Original Paper
Categories

Download Article

How to Cite

1.
Thị Thanh Đông Đỗ, Thị Nhạn N, Kiến Đức N, Hoàng Lâm L, Thị San N, Đình Thanh L. APPLICATION OF SIX SIGMA IN THE EVALUATION AND IMPROVEMENT OF BIOCHEMISTRY TESTING PERFORMANCE AT THONG NHAT HOSPITAL. JHA [Internet]. Vietnam; 2025 Jul. 10 [cited 2025 Jul. 22];1(3):118–124. https://tcsuckhoelaohoa.vn/bvtn/article/view/72 doi: 10.63947/bvtn.v1i3.17
Loading...
Loading citation...

Abstract

The survey aimed to evaluate and improve the performance of testing methods of 10 biochemistry assays on AU5800 (Beckman Coulter) and cobas c503 (Roche Diagnostics) by Sigma metrics. Cross-sectional study. Ten biochemistry assays: glucose, urea, creatinine (Jaffe), cholesterol, triglycerides, AST, GGT, uric acid, HDL-Cho, total bilirubin. Collect internal and external inspection data on 2 devices under the same laboratory conditions, same test materials, at the same time from January to March 2024, calculate Sigma value. Evaluate performance and take improvement action on low-performing tests. On the AU5800 analyzer, 5 out of 10 tests (urea, creatinine, cholesterol, HDL-cholesterol, and total bilirubin) had Sigma values less than 3; 2 out of 10 tests (glucose and AST) had Sigma values between 3 and 6; and 3 out of 10 tests (triglycerides, GGT, and uric acid) had Sigma values greater than 6. On the Cobas c503 analyzer, 1 out of 10 tests (creatinine) had a Sigma value less than 3; 4 out of 10 tests (urea, cholesterol, AST, and total bilirubin) had Sigma values between 3 and 6; and 5 out of 10 tests (glucose, triglycerides, GGT, uric acid, and HDL-cholesterol) had Sigma values greater than 6. The tests with “acceptable to excellent” performance accounted for 50% on the AU5800 analyzer and 90% on the Cobas c503. The creatinine test (Jaffe method) showed “not good” performance on both systems. After improvement by switching to the enzymatic method on the c503 system, its performance changed from “not good” to “excellent.” On the AU5800 system, the performance of the creatinine (Jaffe) test improved from “not good” to “good".

Keywords

Six Sigma analytical performance

References

  1. CLSI (2005). C24-A3 Statistical Quality Control for Quantitative Measurement Procedures: Principles and Definitions. Clinical and Laboratory Standards Institute, Wayne, PA
  2. El Sharkawy Rania, et al. Comparison between Sigma metrics in four accredited Egyptian medical laboratories in some biochemical tests: an initiative towards sigma calculation harmonization. Biochemia Medica. (2018). 233-245 DOI: https://doi.org/10.11613/BM.2018.020711
  3. Feldhammer Matthew, Brown Megan, Colby Jennifer, et al. A survey of Sigma Metrics across three academic medical centers. The Journal of Applied Laboratory Medicine. (2021). 1264-1275. DOI: https://doi.org/10.1093/jalm/jfab028
  4. Ustundag Y, et al. Evaluation of sigma value and quality goal index for brain natriuretic peptide test. Int J Med Biochem.(2020). 178-182. DOI: https://doi.org/10.14744/ijmb.2020.74046
  5. Verma Monica, et al. Assessment of quality control system by sigma metrics and quality goal index ratio: A roadmap towards preparation for NABL. World journal of methodology. (2018). 44. DOI: https://doi.org/10.5662/wjm.v8.i3.44
  6. Westgard James, et al. Basic quality management systems. Madison WI: Westgard QC (2014)

License

© 2025 The Author(s). Published by Journal of Health and Aging.